Measure weight uses a balance or scale to measure the mass of an item. A balance is built using a force restoration mechanism. It measures the force needed to oppose the force applied by the mass placed on the balance. In contrast to a balance which measures mass, a scale measures weight using strain gauges. The mass of a sample can be measured down to a million parts of a gram and up to thousands of grams. Balances with lower maximum loads tend to have higher achievable resolutions, or number of decimal points.
    

Experimental Principles

    Figure 1.1: Procedural overview of a MeasureWeight Experiment. Step 1: The balance is tared zeroing it out. Step 2. The container is transferred to the scale. Step 3. The mass is recorded using the RecordSensor function.

Instrumentation

    Mettler Toledo XP6

    Figure 2.1.1: Items are transferred through an ionizer, which floods the local environment with positive and negative ions. This removes any static electricity that has built up on items and provides a more stable and accurate mass reading.The user must explicitly specify the Micro-Balance using the option Instrument -> Model[Instrument,Balance,"id:54n6evKx08XN"] if so desired.

    Ohaus Pioneer PA124

    Figure 2.2.1: Items are transferred through an ionizer, which floods the local environment with positive and negative ions. This removes any static electricity that has built up on items and provides a more stable and accurate mass reading.

    Ohaus EX6202

    Figure 2.3.1: Items are transferred to the balance to obtain a mass reading.

    D31P60BL

    Figure 2.4.1: Items are transferred to the balance to obtain a mass reading.

Experiment Options

    General

    Instrument

    For each member of 'items', indicates the model or object instrument balance to be used for the weighing.
    Default Value: Automatic
    Default Calculation: Automatic will resolve to an appropriate balance based on the PreferredBalance of the input container's model and the weight of its contents, if known.
    Pattern Description: An object of type or subtype Model[Instrument, Balance] or Object[Instrument, Balance]
    Programmatic Pattern: ObjectP[{Model[Instrument, Balance], Object[Instrument, Balance]}] | Automatic
    Index Matches to: experiment samples

    TransferContainer

    For each member of 'items', indicates if the sample should be transferred to a new container for weighing. Note that if a TransferContainer is specified, the sample will stay in that new container and will not be transferred back to the original container.
    Figure 3.1: The user cannot specify a TransferContainer if CalibrateContainer option is set to True.
    Default Value: Automatic
    Default Calculation: If Automatic, the protocol examines if the existing container holding the sample has a TareWeight associated with it; if no TareWeight is informed, then a new container similar to the existing vessel is determined as TransferContainer for weighing.
    Pattern Description: An object of type or subtype Object[Container, Cuvette], Object[Container, GraduatedCylinder], Object[Container, ReactionVessel], Object[Container, Vessel], Object[Container, ReactionVessel, ElectrochemicalSynthesis], Object[Container, ReactionVessel, Microwave], Object[Container, ReactionVessel, SolidPhaseSynthesis], Object[Container, Vessel, Bag], Object[Container, Vessel, BufferCartridge], Object[Container, Vessel, CrossFlowContainer], Object[Container, Vessel, CrossFlowWashContainer], Object[Container, Vessel, Dialysis], Object[Container, Vessel, Filter], Object[Container, Vessel, GasWashingBottle], Object[Container, Vessel, VolumetricFlask], Model[Container, Cuvette], Model[Container, GraduatedCylinder], Model[Container, ReactionVessel], Model[Container, Vessel], Model[Container, ReactionVessel, ElectrochemicalSynthesis], Model[Container, ReactionVessel, Microwave], Model[Container, ReactionVessel, SolidPhaseSynthesis], Model[Container, Vessel, Bag], Model[Container, Vessel, BufferCartridge], Model[Container, Vessel, CrossFlowContainer], Model[Container, Vessel, CrossFlowWashContainer], Model[Container, Vessel, Dialysis], Model[Container, Vessel, Filter], Model[Container, Vessel, GasWashingBottle], or Model[Container, Vessel, VolumetricFlask] or a prepared sample or Null.
    Programmatic Pattern: ((ObjectP[Flatten[Join[{MeasureWeightContainerTypes, MeasureWeightModelContainerTypes}]]] | _String) | Automatic) | Null
    Index Matches to: experiment samples

    CalibrateContainer

    For each member of 'items, indicates if the TareWeight of the container will be measured instead of the weight of the sample. If True, the weight of any sample in the container must be known, unless the container is empty.
    Figure 3.2: The user cannot specify a TransferContainer if CalibrateContainer option is set to True.
    Default Value: Automatic
    Default Calculation: Automatic will resolve to False if Mass (in Object[Sample]) is unknown or True if Mass is already informed.
    Pattern Description: True or False.
    Programmatic Pattern: BooleanP | Automatic
    Index Matches to: experiment samples

    NumberOfReplicates

    The number of times to repeat the experiment on each provided 'item'. If specified, the recorded replicate measurements will be averaged for determining the final Mass of the sample(s). Note that specifying NumberOfReplicates is identical to providing duplicated input.
    Default Value: Null
    Pattern Description: Greater than or equal to 2 and less than or equal to 10 in increments of 1 or Null.
    Programmatic Pattern: RangeP[2, 10, 1] | Null

    Post Experiment

    SamplesInStorageCondition

    The non-default conditions under which the SamplesIn of this experiment should be stored after the protocol is completed. If left unset, SamplesIn will be stored according to their current StorageCondition.
    Default Value: Null
    Pattern Description: {AmbientStorage, Refrigerator, Freezer, DeepFreezer, CryogenicStorage, YeastIncubation, BacteriaIncubation, MammalianIncubation, TissueCultureCellsIncubation, MicrobialCellsIncubation, MicrobialCellsShakingIncubation, YeastCellsIncubation, YeastCellsShakingIncubation, ViralIncubation, AcceleratedTesting, IntermediateTesting, LongTermTesting, UVVisLightTesting} or Disposal or Null.
    Programmatic Pattern: (Alternatives[SampleStorageTypeP | Disposal]) | Null
    Index Matches to: experiment samples

    SamplesOutStorageCondition

    The non-default conditions under which any new samples generated by this experiment should be stored after the protocol is completed. If left unset, the new samples will be stored according to their Models' DefaultStorageCondition.
    Default Value: Null
    Pattern Description: {AmbientStorage, Refrigerator, Freezer, DeepFreezer, CryogenicStorage, YeastIncubation, BacteriaIncubation, MammalianIncubation, TissueCultureCellsIncubation, MicrobialCellsIncubation, MicrobialCellsShakingIncubation, YeastCellsIncubation, YeastCellsShakingIncubation, ViralIncubation, AcceleratedTesting, IntermediateTesting, LongTermTesting, UVVisLightTesting} or Disposal or Null.
    Programmatic Pattern: (Alternatives[SampleStorageTypeP | Disposal]) | Null
    Index Matches to: experiment samples

Sample Prep Options

    Aliquoting

    Aliquot

    Indicates if aliquots should be taken from the SamplesIn and transferred into new AliquotSamples used in lieu of the SamplesIn for the experiment. Note that if NumberOfReplicates is specified this indicates that the input samples will also be aliquoted that number of times. Note that Aliquoting (if specified) occurs after any Sample Preparation (if specified).
    Default Value: Automatic
    Pattern Description: True or False.
    Programmatic Pattern: BooleanP | Automatic
    Index Matches to: experiment samples

    AliquotAmount

    The amount of a sample that should be transferred from the input samples into aliquots.
    Default Value: Automatic
    Default Calculation: Automatically set as the smaller between the current sample volume and the maximum volume of the destination container if a liquid, or the current Mass or Count if a solid or counted item, respectively.
    Pattern Description: All or Count or Count or Mass or Volume or Null.
    Programmatic Pattern: ((RangeP[1*Microliter, 20*Liter] | RangeP[1*Milligram, 20*Kilogram] | GreaterP[0*Unit, 1*Unit] | GreaterP[0., 1.] | All) | Automatic) | Null
    Index Matches to: experiment samples

    TargetConcentration

    The desired final concentration of analyte in the AliquotSamples after dilution of aliquots of SamplesIn with the ConcentratedBuffer and BufferDiluent which should be used in lieu of the SamplesIn for the experiment.
    Default Value: Automatic
    Default Calculation: Automatically calculated based on aliquot and buffer volumes.
    Pattern Description: Greater than 0 molar or greater than 0 grams per liter or Null.
    Programmatic Pattern: ((GreaterP[0*Molar] | GreaterP[0*(Gram/Liter)]) | Automatic) | Null
    Index Matches to: experiment samples

    TargetConcentrationAnalyte

    The substance whose final concentration is attained with the TargetConcentration option.
    Default Value: Automatic
    Default Calculation: Automatically set to the first value in the Analytes field of the input sample, or, if not populated, to the first analyte in the Composition field of the input sample, or if none exist, the first identity model of any kind in the Composition field.
    Pattern Description: An object of type or subtype Model[Molecule], Model[Molecule, cDNA], Model[Molecule, Oligomer], Model[Molecule, Transcript], Model[Molecule, Protein], Model[Molecule, Protein, Antibody], Model[Molecule, Carbohydrate], Model[Molecule, Polymer], Model[Resin], Model[Resin, SolidPhaseSupport], Model[Lysate], Model[ProprietaryFormulation], Model[Virus], Model[Cell], Model[Cell, Mammalian], Model[Cell, Bacteria], Model[Cell, Yeast], Model[Tissue], Model[Material], or Model[Species] or Null.
    Programmatic Pattern: (ObjectP[IdentityModelTypes] | Automatic) | Null
    Index Matches to: experiment samples

    AssayVolume

    The desired total volume of the aliquoted sample plus dilution buffer.
    Default Value: Automatic
    Default Calculation: Automatically determined based on Volume and TargetConcentration option values.
    Pattern Description: Greater than or equal to 1 microliter and less than or equal to 20 liters or Null.
    Programmatic Pattern: (RangeP[1*Microliter, 20*Liter] | Automatic) | Null
    Index Matches to: experiment samples

    ConcentratedBuffer

    The concentrated buffer which should be diluted by the BufferDilutionFactor in the final solution (i.e., the combination of the sample, ConcentratedBuffer, and BufferDiluent). The ConcentratedBuffer and BufferDiluent will be combined and then mixed with the sample, where the combined volume of these buffers is the difference between the AliquotAmount and the total AssayVolume.
    Default Value: Automatic
    Pattern Description: An object of type or subtype Model[Sample] or Object[Sample] or a prepared sample or Null.
    Programmatic Pattern: ((ObjectP[{Model[Sample], Object[Sample]}] | _String) | Automatic) | Null
    Index Matches to: experiment samples

    BufferDilutionFactor

    The dilution factor by which the concentrated buffer should be diluted in the final solution (i.e., the combination of the sample, ConcentratedBuffer, and BufferDiluent). The ConcentratedBuffer and BufferDiluent will be combined and then mixed with the sample, where the combined volume of these buffers is the difference between the AliquotAmount and the total AssayVolume.
    Default Value: Automatic
    Default Calculation: If ConcentratedBuffer is specified, automatically set to the ConcentrationFactor of that sample; otherwise, set to Null.
    Pattern Description: Greater than or equal to 1 or Null.
    Programmatic Pattern: (GreaterEqualP[1] | Automatic) | Null
    Index Matches to: experiment samples

    BufferDiluent

    The buffer used to dilute the aliquot sample such that ConcentratedBuffer is diluted by BufferDilutionFactor in the final solution. The ConcentratedBuffer and BufferDiluent will be combined and then mixed with the sample, where the combined volume of these buffers is the difference between the AliquotAmount and the total AssayVolume.
    Default Value: Automatic
    Default Calculation: Automatically resolves to Model[Sample, "Milli-Q water"] if ConcentratedBuffer is specified; otherwise, resolves to Null.
    Pattern Description: An object of type or subtype Model[Sample] or Object[Sample] or a prepared sample or Null.
    Programmatic Pattern: ((ObjectP[{Model[Sample], Object[Sample]}] | _String) | Automatic) | Null
    Index Matches to: experiment samples

    AssayBuffer

    The buffer that should be added to any aliquots requiring dilution, where the volume of this buffer added is the difference between the AliquotAmount and the total AssayVolume.
    Default Value: Automatic
    Default Calculation: Automatically resolves to Model[Sample, "Milli-Q water"] if ConcentratedBuffer is not specified; otherwise, resolves to Null.
    Pattern Description: An object of type or subtype Model[Sample] or Object[Sample] or a prepared sample or Null.
    Programmatic Pattern: ((ObjectP[{Model[Sample], Object[Sample]}] | _String) | Automatic) | Null
    Index Matches to: experiment samples

    AliquotSampleStorageCondition

    The non-default conditions under which any aliquot samples generated by this experiment should be stored after the protocol is completed.
    Default Value: Automatic
    Pattern Description: {AmbientStorage, Refrigerator, Freezer, DeepFreezer, CryogenicStorage, YeastIncubation, BacteriaIncubation, MammalianIncubation, TissueCultureCellsIncubation, MicrobialCellsIncubation, MicrobialCellsShakingIncubation, YeastCellsIncubation, YeastCellsShakingIncubation, ViralIncubation, AcceleratedTesting, IntermediateTesting, LongTermTesting, UVVisLightTesting} or Disposal or Null.
    Programmatic Pattern: ((SampleStorageTypeP | Disposal) | Automatic) | Null
    Index Matches to: experiment samples

    DestinationWell

    The desired position in the corresponding AliquotContainer in which the aliquot samples will be placed.
    Default Value: Automatic
    Default Calculation: Automatically resolves to A1 in containers with only one position. For plates, fills wells in the order provided by the function AllWells.
    Pattern Description: Any well from A1 to H12 or list of one or more any well from A1 to H12 or any well from A1 to H12 entries or Null.
    Programmatic Pattern: ((WellPositionP | {((Automatic | Null) | WellPositionP)..}) | Automatic) | Null

    AliquotContainer

    The desired type of container that should be used to prepare and house the aliquot samples, with indices indicating grouping of samples in the same plates, if desired. This option will resolve to be the length of the SamplesIn * NumberOfReplicates.
    Default Value: Automatic
    Default Calculation: Automatically set as the PreferredContainer for the AssayVolume of the sample. For plates, attempts to fill all wells of a single plate with the same model before aliquoting into the next.
    Pattern Description: An object of type or subtype Model[Container] or Object[Container] or a prepared sample or Automatic or Null or {Index, Container} or list of one or more an object of type or subtype Model[Container] or Object[Container] or a prepared sample or Automatic or Null entries or list of one or more Automatic or Null or {Index, Container} entries or Null.
    Programmatic Pattern: (((ObjectP[{Model[Container], Object[Container]}] | _String) | (Automatic | Null) | {GreaterEqualP[1, 1] | (Automatic | Null), (ObjectP[{Model[Container], Object[Container]}] | _String) | (Automatic | Null)} | {((ObjectP[{Model[Container], Object[Container]}] | _String) | (Automatic | Null))..} | {({GreaterEqualP[1, 1] | (Automatic | Null), (ObjectP[{Model[Container], Object[Container]}] | _String) | (Automatic | Null)} | (Automatic | Null))..}) | Automatic) | Null

    AliquotPreparation

    Indicates the desired scale at which liquid handling used to generate aliquots will occur.
    Default Value: Automatic
    Default Calculation: Automatic resolution will occur based on manipulation volumes and container types.
    Pattern Description: Manual or Robotic or Null.
    Programmatic Pattern: (PreparationMethodP | Automatic) | Null

    ConsolidateAliquots

    Indicates if identical aliquots should be prepared in the same container/position.
    Default Value: Automatic
    Pattern Description: True or False or Null.
    Programmatic Pattern: (BooleanP | Automatic) | Null

    Sample Preparation

    PreparatoryUnitOperations

    Specifies a sequence of transferring, aliquoting, consolidating, or mixing of new or existing samples before the main experiment. These prepared samples can be used in the main experiment by referencing their defined name. For more information, please reference the documentation for ExperimentSampleManipulation.
    Default Value: Null
    Pattern Description: List of one or more unit Operation ManualSamplePreparation or RoboticSamplePreparation or unit Operation must match SamplePreparationP entries or Null.
    Programmatic Pattern: {((ManualSamplePreparationMethodP | RoboticSamplePreparationMethodP) | SamplePreparationP)..} | Null

    PreparatoryPrimitives

    Specifies a sequence of transferring, aliquoting, consolidating, or mixing of new or existing samples before the main experiment. These prepared samples can be used in the main experiment by referencing their defined name. For more information, please reference the documentation for ExperimentSampleManipulation.
    Default Value: Null
    Pattern Description: List of one or more a primitive with head Define, Transfer, Mix, Aliquot, Consolidation, FillToVolume, Incubate, Filter, Wait, Centrifuge, or Resuspend entries or Null.
    Programmatic Pattern: {SampleManipulationP..} | Null

    Preparatory Incubation

    Incubate

    Indicates if the SamplesIn should be incubated at a fixed temperature prior to starting the experiment or any aliquoting. Sample Preparation occurs in the order of Incubation, Centrifugation, Filtration, and then Aliquoting (if specified).
    Default Value: Automatic
    Default Calculation: Resolves to True if any of the corresponding Incubation options are set. Otherwise, resolves to False.
    Pattern Description: True or False.
    Programmatic Pattern: BooleanP | Automatic
    Index Matches to: experiment samples

    IncubationTemperature

    Temperature at which the SamplesIn should be incubated for the duration of the IncubationTime prior to starting the experiment.
    Default Value: Automatic
    Pattern Description: Ambient or greater than or equal to -20 degrees Celsius and less than or equal to 500 degrees Celsius or Null.
    Programmatic Pattern: ((Ambient | RangeP[$MinIncubationTemperature, $MaxIncubationTemperature]) | Automatic) | Null
    Index Matches to: experiment samples

    IncubationTime

    Duration for which SamplesIn should be incubated at the IncubationTemperature, prior to starting the experiment.
    Default Value: Automatic
    Pattern Description: Greater than or equal to 1 minute and less than or equal to 72 hours or Null.
    Programmatic Pattern: (RangeP[1*Minute, $MaxExperimentTime] | Automatic) | Null
    Index Matches to: experiment samples

    Mix

    Indicates if this sample should be mixed while incubated, prior to starting the experiment.
    Default Value: Automatic
    Default Calculation: Automatically resolves to True if any Mix related options are set. Otherwise, resolves to False.
    Pattern Description: True or False or Null.
    Programmatic Pattern: (BooleanP | Automatic) | Null
    Index Matches to: experiment samples

    MixType

    Indicates the style of motion used to mix the sample, prior to starting the experiment.
    Default Value: Automatic
    Default Calculation: Automatically resolves based on the container of the sample and the Mix option.
    Pattern Description: Roll, Vortex, Sonicate, Pipette, Invert, Stir, Shake, Homogenize, Swirl, Disrupt, or Nutate or Null.
    Programmatic Pattern: (MixTypeP | Automatic) | Null
    Index Matches to: experiment samples

    MixUntilDissolved

    Indicates if the mix should be continued up to the MaxIncubationTime or MaxNumberOfMixes (chosen according to the mix Type), in an attempt dissolve any solute. Any mixing/incubation will occur prior to starting the experiment.
    Default Value: Automatic
    Default Calculation: Automatically resolves to True if MaxIncubationTime or MaxNumberOfMixes is set.
    Pattern Description: True or False or Null.
    Programmatic Pattern: (BooleanP | Automatic) | Null
    Index Matches to: experiment samples

    MaxIncubationTime

    Maximum duration of time for which the samples will be mixed while incubated in an attempt to dissolve any solute, if the MixUntilDissolved option is chosen. This occurs prior to starting the experiment.
    Default Value: Automatic
    Default Calculation: Automatically resolves based on MixType, MixUntilDissolved, and the container of the given sample.
    Pattern Description: Greater than or equal to 1 minute and less than or equal to 72 hours or Null.
    Programmatic Pattern: (RangeP[1*Minute, $MaxExperimentTime] | Automatic) | Null
    Index Matches to: experiment samples

    IncubationInstrument

    The instrument used to perform the Mix and/or Incubation, prior to starting the experiment.
    Default Value: Automatic
    Default Calculation: Automatically resolves based on the options Mix, Temperature, MixType and container of the sample.
    Pattern Description: An object of type or subtype Model[Instrument, Roller], Model[Instrument, OverheadStirrer], Model[Instrument, Vortex], Model[Instrument, Shaker], Model[Instrument, BottleRoller], Model[Instrument, Roller], Model[Instrument, Sonicator], Model[Instrument, HeatBlock], Model[Instrument, Homogenizer], Model[Instrument, Disruptor], Model[Instrument, Nutator], Model[Instrument, Thermocycler], Model[Instrument, EnvironmentalChamber], Model[Instrument, Pipette], Object[Instrument, Roller], Object[Instrument, OverheadStirrer], Object[Instrument, Vortex], Object[Instrument, Shaker], Object[Instrument, BottleRoller], Object[Instrument, Roller], Object[Instrument, Sonicator], Object[Instrument, HeatBlock], Object[Instrument, Homogenizer], Object[Instrument, Disruptor], Object[Instrument, Nutator], Object[Instrument, Thermocycler], Object[Instrument, EnvironmentalChamber], or Object[Instrument, Pipette] or Null.
    Programmatic Pattern: (ObjectP[Join[MixInstrumentModels, MixInstrumentObjects]] | Automatic) | Null
    Index Matches to: experiment samples

    AnnealingTime

    Minimum duration for which the SamplesIn should remain in the incubator allowing the system to settle to room temperature after the IncubationTime has passed but prior to starting the experiment.
    Default Value: Automatic
    Pattern Description: Greater than or equal to 0 minutes and less than or equal to 72 hours or Null.
    Programmatic Pattern: (RangeP[0*Minute, $MaxExperimentTime] | Automatic) | Null
    Index Matches to: experiment samples

    IncubateAliquotContainer

    The desired type of container that should be used to prepare and house the incubation samples which should be used in lieu of the SamplesIn for the experiment.
    Default Value: Automatic
    Pattern Description: An object of type or subtype Model[Container] or {Index, Container} or Null.
    Programmatic Pattern: ((ObjectP[Model[Container]] | {GreaterEqualP[1, 1] | (Automatic | Null), (ObjectP[{Model[Container], Object[Container]}] | _String) | Automatic}) | Automatic) | Null
    Index Matches to: experiment samples

    IncubateAliquotDestinationWell

    The desired position in the corresponding IncubateAliquotContainer in which the aliquot samples will be placed.
    Default Value: Automatic
    Default Calculation: Automatically resolves to A1 in containers with only one position. For plates, fills wells in the order provided by the function AllWells.
    Pattern Description: Any well from A1 to H12 or Null.
    Programmatic Pattern: (WellPositionP | Automatic) | Null
    Index Matches to: experiment samples

    IncubateAliquot

    The amount of each sample that should be transferred from the SamplesIn into the IncubateAliquotContainer when performing an aliquot before incubation.
    Default Value: Automatic
    Default Calculation: Automatically set as the smaller between the current sample volume and the maximum volume of the destination container.
    Pattern Description: All or greater than or equal to 1 microliter and less than or equal to 20 liters or Null.
    Programmatic Pattern: ((RangeP[1*Microliter, 20*Liter] | All) | Automatic) | Null
    Index Matches to: experiment samples

    Preparatory Centrifugation

    Centrifuge

    Indicates if the SamplesIn should be centrifuged prior to starting the experiment or any aliquoting. Sample Preparation occurs in the order of Incubation, Centrifugation, Filtration, and then Aliquoting (if specified).
    Default Value: Automatic
    Default Calculation: Resolves to True if any of the corresponding Centrifuge options are set. Otherwise, resolves to False.
    Pattern Description: True or False.
    Programmatic Pattern: BooleanP | Automatic
    Index Matches to: experiment samples

    CentrifugeInstrument

    The centrifuge that will be used to spin the provided samples prior to starting the experiment.
    Default Value: Automatic
    Pattern Description: An object of type or subtype Model[Instrument, Centrifuge] or Object[Instrument, Centrifuge] or Null.
    Programmatic Pattern: (ObjectP[{Model[Instrument, Centrifuge], Object[Instrument, Centrifuge]}] | Automatic) | Null
    Index Matches to: experiment samples

    CentrifugeIntensity

    The rotational speed or the force that will be applied to the samples by centrifugation prior to starting the experiment.
    Default Value: Automatic
    Pattern Description: Greater than 0 revolutions per minute or greater than 0 standard accelerations due to gravity on the surface of the earth or Null.
    Programmatic Pattern: ((GreaterP[0*RPM] | GreaterP[0*GravitationalAcceleration]) | Automatic) | Null
    Index Matches to: experiment samples

    CentrifugeTime

    The amount of time for which the SamplesIn should be centrifuged prior to starting the experiment.
    Default Value: Automatic
    Pattern Description: Greater than 0 minutes or Null.
    Programmatic Pattern: (GreaterP[0*Minute] | Automatic) | Null
    Index Matches to: experiment samples

    CentrifugeTemperature

    The temperature at which the centrifuge chamber should be held while the samples are being centrifuged prior to starting the experiment.
    Default Value: Automatic
    Pattern Description: Ambient or greater than or equal to -10 degrees Celsius and less than or equal to 40 degrees Celsius or Null.
    Programmatic Pattern: ((Ambient | RangeP[-10*Celsius, 40*Celsius]) | Automatic) | Null
    Index Matches to: experiment samples

    CentrifugeAliquotContainer

    The desired type of container that should be used to prepare and house the centrifuge samples which should be used in lieu of the SamplesIn for the experiment.
    Default Value: Automatic
    Pattern Description: An object of type or subtype Model[Container] or {Index, Container} or Null.
    Programmatic Pattern: ((ObjectP[Model[Container]] | {GreaterEqualP[1, 1] | (Automatic | Null), (ObjectP[{Model[Container], Object[Container]}] | _String) | Automatic}) | Automatic) | Null
    Index Matches to: experiment samples

    CentrifugeAliquotDestinationWell

    The desired position in the corresponding AliquotContainer in which the aliquot samples will be placed.
    Default Value: Automatic
    Default Calculation: Automatically resolves to A1 in containers with only one position. For plates, fills wells in the order provided by the function AllWells.
    Pattern Description: Any well from A1 to H12 or Null.
    Programmatic Pattern: (WellPositionP | Automatic) | Null
    Index Matches to: experiment samples

    CentrifugeAliquot

    The amount of each sample that should be transferred from the SamplesIn into the CentrifugeAliquotContainer when performing an aliquot before centrifugation.
    Default Value: Automatic
    Default Calculation: Automatically set as the smaller between the current sample volume and the maximum volume of the destination container.
    Pattern Description: All or greater than or equal to 1 microliter and less than or equal to 20 liters or Null.
    Programmatic Pattern: ((RangeP[1*Microliter, 20*Liter] | All) | Automatic) | Null
    Index Matches to: experiment samples

    Preparatory Filtering

    Filtration

    Indicates if the SamplesIn should be filter prior to starting the experiment or any aliquoting. Sample Preparation occurs in the order of Incubation, Centrifugation, Filtration, and then Aliquoting (if specified).
    Default Value: Automatic
    Default Calculation: Resolves to True if any of the corresponding Filter options are set. Otherwise, resolves to False.
    Pattern Description: True or False.
    Programmatic Pattern: BooleanP | Automatic
    Index Matches to: experiment samples

    FiltrationType

    The type of filtration method that should be used to perform the filtration.
    Default Value: Automatic
    Default Calculation: Will automatically resolve to a filtration type appropriate for the volume of sample being filtered.
    Pattern Description: PeristalticPump, Centrifuge, Vacuum, Syringe, or AirPressure or Null.
    Programmatic Pattern: (FiltrationTypeP | Automatic) | Null
    Index Matches to: experiment samples

    FilterInstrument

    The instrument that should be used to perform the filtration.
    Default Value: Automatic
    Default Calculation: Will automatically resolved to an instrument appropriate for the filtration type.
    Pattern Description: An object of type or subtype Model[Instrument, FilterBlock], Object[Instrument, FilterBlock], Model[Instrument, PeristalticPump], Object[Instrument, PeristalticPump], Model[Instrument, VacuumPump], Object[Instrument, VacuumPump], Model[Instrument, Centrifuge], Object[Instrument, Centrifuge], Model[Instrument, SyringePump], or Object[Instrument, SyringePump] or Null.
    Programmatic Pattern: (ObjectP[{Model[Instrument, FilterBlock], Object[Instrument, FilterBlock], Model[Instrument, PeristalticPump], Object[Instrument, PeristalticPump], Model[Instrument, VacuumPump], Object[Instrument, VacuumPump], Model[Instrument, Centrifuge], Object[Instrument, Centrifuge], Model[Instrument, SyringePump], Object[Instrument, SyringePump]}] | Automatic) | Null
    Index Matches to: experiment samples

    Filter

    The filter that should be used to remove impurities from the SamplesIn prior to starting the experiment.
    Default Value: Automatic
    Default Calculation: Will automatically resolve to a filter appropriate for the filtration type and instrument.
    Pattern Description: An object of type or subtype Model[Container, Plate, Filter], Model[Container, Vessel, Filter], or Model[Item, Filter] or Null.
    Programmatic Pattern: (ObjectP[{Model[Container, Plate, Filter], Model[Container, Vessel, Filter], Model[Item, Filter]}] | Automatic) | Null
    Index Matches to: experiment samples

    FilterMaterial

    The membrane material of the filter that should be used to remove impurities from the SamplesIn prior to starting the experiment.
    Default Value: Automatic
    Default Calculation: Resolves to an appropriate filter material for the given sample is Filtration is set to True.
    Pattern Description: Cellulose, Cotton, Polyethylene, PTFE, Nylon, PES, PLUS, PVDF, GlassFiber, GHP, UHMWPE, EPDM, DuraporePVDF, GxF, ZebaDesaltingResin, NickelResin, Silica, or HLB or Null.
    Programmatic Pattern: (FilterMembraneMaterialP | Automatic) | Null
    Index Matches to: experiment samples

    PrefilterMaterial

    The material from which the prefilter filtration membrane should be made of to remove impurities from the SamplesIn prior to starting the experiment.
    Default Value: Automatic
    Default Calculation: By default, no prefiltration is performed on samples, even when Filter->True.
    Pattern Description: Cellulose, Cotton, Polyethylene, PTFE, Nylon, PES, PLUS, PVDF, GlassFiber, GHP, UHMWPE, EPDM, DuraporePVDF, GxF, ZebaDesaltingResin, NickelResin, Silica, or HLB or Null.
    Programmatic Pattern: (FilterMembraneMaterialP | Automatic) | Null
    Index Matches to: experiment samples

    FilterPoreSize

    The pore size of the filter that should be used when removing impurities from the SamplesIn prior to starting the experiment.
    Default Value: Automatic
    Default Calculation: Resolves to an appropriate filter pore size for the given sample is Filtration is set to True.
    Pattern Description: 0.008 micrometers, 0.1 micrometers, 0.22 micrometers, 0.45 micrometers, 1. micrometer, 1.1 micrometers, 2.5 micrometers, 6. micrometers, 20. micrometers, 30. micrometers, or 100. micrometers or Null.
    Programmatic Pattern: (FilterSizeP | Automatic) | Null
    Index Matches to: experiment samples

    PrefilterPoreSize

    The pore size of the filter; all particles larger than this should be removed during the filtration.
    Default Value: Automatic
    Default Calculation: By default, no prefiltration is performed on samples, even when Filter->True.
    Pattern Description: 0.008 micrometers, 0.1 micrometers, 0.22 micrometers, 0.45 micrometers, 1. micrometer, 1.1 micrometers, 2.5 micrometers, 6. micrometers, 20. micrometers, 30. micrometers, or 100. micrometers or Null.
    Programmatic Pattern: (FilterSizeP | Automatic) | Null
    Index Matches to: experiment samples

    FilterSyringe

    The syringe used to force that sample through a filter.
    Default Value: Automatic
    Default Calculation: Resolves to an syringe appropriate to the volume of sample being filtered, if Filtration is set to True.
    Pattern Description: An object of type or subtype Model[Container, Syringe] or Object[Container, Syringe] or a prepared sample or Null.
    Programmatic Pattern: ((ObjectP[{Model[Container, Syringe], Object[Container, Syringe]}] | _String) | Automatic) | Null
    Index Matches to: experiment samples

    FilterHousing

    The filter housing that should be used to hold the filter membrane when filtration is performed using a standalone filter membrane.
    Default Value: Automatic
    Default Calculation: Resolve to an housing capable of holding the size of the membrane being used, if filter with Membrane FilterType is being used and Filtration is set to True.
    Pattern Description: An object of type or subtype Model[Instrument, FilterHousing], Object[Instrument, FilterHousing], Model[Instrument, FilterBlock], or Object[Instrument, FilterBlock] or Null.
    Programmatic Pattern: (ObjectP[{Model[Instrument, FilterHousing], Object[Instrument, FilterHousing], Model[Instrument, FilterBlock], Object[Instrument, FilterBlock]}] | Automatic) | Null
    Index Matches to: experiment samples

    FilterIntensity

    The rotational speed or force at which the samples will be centrifuged during filtration.
    Default Value: Automatic
    Default Calculation: Will automatically resolve to 2000 GravitationalAcceleration if FiltrationType is Centrifuge and Filtration is True.
    Pattern Description: Greater than 0 revolutions per minute or greater than 0 standard accelerations due to gravity on the surface of the earth or Null.
    Programmatic Pattern: ((GreaterP[0*RPM] | GreaterP[0*GravitationalAcceleration]) | Automatic) | Null
    Index Matches to: experiment samples

    FilterTime

    The amount of time for which the samples will be centrifuged during filtration.
    Default Value: Automatic
    Default Calculation: Will automatically resolve to 5 Minute if FiltrationType is Centrifuge and Filtration is True.
    Pattern Description: Greater than 0 minutes or Null.
    Programmatic Pattern: (GreaterP[0*Minute] | Automatic) | Null
    Index Matches to: experiment samples

    FilterTemperature

    The temperature at which the centrifuge chamber will be held while the samples are being centrifuged during filtration.
    Default Value: Automatic
    Default Calculation: Will automatically resolve to 22 Celsius if FiltrationType is Centrifuge and Filtration is True.
    Pattern Description: Greater than or equal to 4 degrees Celsius or Null.
    Programmatic Pattern: ((Alternatives[GreaterEqualP[4*Celsius]]) | Automatic) | Null
    Index Matches to: experiment samples

    FilterContainerOut

    The desired container filtered samples should be produced in or transferred into by the end of filtration, with indices indicating grouping of samples in the same plates, if desired.
    Default Value: Automatic
    Default Calculation: Automatically set as the PreferredContainer for the Volume of the sample. For plates, attempts to fill all wells of a single plate with the same model before using another one.
    Pattern Description: An object of type or subtype Model[Container] or Object[Container] or a prepared sample or {Index, Container} or Null.
    Programmatic Pattern: (((ObjectP[{Model[Container], Object[Container]}] | _String) | {GreaterEqualP[1, 1] | Automatic, (ObjectP[{Model[Container], Object[Container]}] | _String) | Automatic}) | Automatic) | Null
    Index Matches to: experiment samples

    FilterAliquotDestinationWell

    The desired position in the corresponding AliquotContainer in which the aliquot samples will be placed.
    Default Value: Automatic
    Default Calculation: Automatically resolves to A1 in containers with only one position. For plates, fills wells in the order provided by the function AllWells.
    Pattern Description: Any well from A1 to H12 or Null.
    Programmatic Pattern: (WellPositionP | Automatic) | Null
    Index Matches to: experiment samples

    FilterAliquotContainer

    The desired type of container that should be used to prepare and house the filter samples which should be used in lieu of the SamplesIn for the experiment.
    Default Value: Automatic
    Pattern Description: An object of type or subtype Model[Container] or {Index, Container} or Null.
    Programmatic Pattern: ((ObjectP[Model[Container]] | {GreaterEqualP[1, 1] | (Automatic | Null), (ObjectP[{Model[Container], Object[Container]}] | _String) | Automatic}) | Automatic) | Null
    Index Matches to: experiment samples

    FilterAliquot

    The amount of each sample that should be transferred from the SamplesIn into the FilterAliquotContainer when performing an aliquot before filtration.
    Default Value: Automatic
    Default Calculation: Automatically set as the smaller between the current sample volume and the maximum volume of the destination container.
    Pattern Description: All or greater than or equal to 1 microliter and less than or equal to 20 liters or Null.
    Programmatic Pattern: ((RangeP[1*Microliter, 20*Liter] | All) | Automatic) | Null
    Index Matches to: experiment samples

    FilterSterile

    Indicates if the filtration of the samples should be done in a sterile environment.
    Default Value: Automatic
    Default Calculation: Resolve to False if Filtration is indicated. If sterile filtration is desired, this option must manually be set to True.
    Pattern Description: True or False or Null.
    Programmatic Pattern: (BooleanP | Automatic) | Null
    Index Matches to: experiment samples

Protocol Options

    Organizational Information

    Template

    A template protocol whose methodology should be reproduced in running this experiment. Option values will be inherited from the template protocol, but can be individually overridden by directly specifying values for those options to this Experiment function.
    Default Value: Null
    Pattern Description: An object of type or subtype Object[Protocol] or an object of type or subtype of Object[Protocol] with UnresolvedOptions, ResolvedOptions specified or Null.
    Programmatic Pattern: (ObjectP[Object[Protocol]] | FieldReferenceP[Object[Protocol], {UnresolvedOptions, ResolvedOptions}]) | Null

    Name

    A object name which should be used to refer to the output object in lieu of an automatically generated ID number.
    Default Value: Null
    Pattern Description: A string or Null.
    Programmatic Pattern: _String | Null

    Post Experiment

    ImageSample

    Indicates if any samples that are modified in the course of the experiment should be freshly imaged after running the experiment. Please note that public samples are imaged regardless of the value of this option.
    Default Value: Automatic
    Pattern Description: True or False or Null.
    Programmatic Pattern: (BooleanP | Automatic) | Null

Example Calls

    Transfer Container

    Specify a model container to transfer your sample into before weighing:

    Tare Weight of Container

    Determine the tare weight of your container:

Preferred Input Containers

    The analytical balance is used to measure samples up to 120g
    The macro balance is used to measure samples up to 6Kg
    The bulk balance is used to measure samples greater than 6kg

Warnings and Errors

    Messages  (17)

    ConflictingOptions  (1)

    If CalibrateContainer is specified to True, we cannot have TransferContainer specified as well:

    ContainerIncompatibleWithBalance  (1)

    Instrument can only be specified to a Microbalance if the containers are of the model {Model[Container,Vessel,"id:AEqRl954GG7a"],Model[Container,Vessel,"id:7X104vK9ZZNJ"]} or a TransferContainer is specified:

    DiscardedSamples  (1)

    If the given input samples are discarded, they cannot be weighed:

    ImmobileSamples  (1)

    Immobile samples cannot be measured because there are fixed in place:

    IncompatibleContainerType  (1)

    If the input containers or the containers of the input samples are not of the type MeasureWeightContainerP, they cannot be weighed:

    InputContainsTemporalLinks  (1)

    A Warning is thrown if any inputs or options contain temporal links:

    MissingContainer  (1)

    If the given input samples do not have a Container, they can't be weighed:

    NoContentsToBeTransferred  (1)

    Cannot specify a TransferContainer if the input containers do not have a content:

    SampleMassMayBeInaccurate  (1)

    When CalibrateContainer is set to True, it is assumed that the sample's mass is trustworthy (i.e. has not been manipulated by any other protocol after the latest mass entry has been recorded):

    SampleMassUnknown  (1)

    We can only measure the tareweight of the input container (CalibrateContainer -> True) if the container is either empty or it contains a sample whose weight is known:

    SamplePrepOnEmptyContainer  (1)

    If the given input container is empty, SamplePrep options cannot be specified:

    TareWeightNeeded  (1)

    TransferContainer can only be specified to Null if the input container's tareweight is known (or is empty):

    TransferContainerNotEmpty  (1)

    Specified TransferContainer has to be empty:

    UnsuitableBalance  (1)

    Specified Instrument needs to be identical to the balance that would be recommended for that particular sample:

    UnsuitableMicroBalance  (2)

    Instrument can only be specified to a Microbalance if the weight of the sample is trustworthy:

    Instrument can only be specified to a Microbalance if the weight of the sample to be weighed is below the weighing limit of the micro balance (4.8 Gram):

    VentilatedSample  (1)

    It is not currently possible to weigh samples which require a transfer container and must be handled in a fume hood: